Precision, repeatability and accuracy of Optotrak® optical motion tracking systems

##article.authors##

DOI:

https://doi.org/10.31224/osf.io/mcr5x

Keywords:

biomechanics, engrxiv, Preprint, validation

Abstract

In the field of biomechanics, optical motion tracking systems are commonly used to record human motion and assist in surgical navigation. Recently, motion tracking systems have been used to track implant and bone motion on a micron-level. The present study evaluated four different Optotrak® motion tracking systems to determine the precision, repeatability and accuracy under static testing conditions. The distance between the camera systems and the rigid body, as well as the tilt angle of the rigid body, did affect the resulting precision, repeatability and accuracy of the camera systems. The precision and repeatability, calculated as the within-trial and between-trial standard deviations, respectively, were less than 30 µm; with some configurations producing precision and repeatability less than 1 µm. The accuracy was less than 0.53% of the total displacement for the in-plane motion and less than 1.56% of the total displacement for the out-of-plane motion.

Downloads

Download data is not yet available.

Posted

2016-07-27